@article{oai:tsukuba.repo.nii.ac.jp:00051904, author = {加納, 英明 and KANOU, Hideaki and 野口, 雅之 and NOGUCHI, Masayuki and Maruyama, Takumi and Kano, Junko and Oka, Yuki and Kaneta, Daiki and Guerenne, Tiffany and Leproux, Philippe and Couderc, Vincent}, issue = {5}, journal = {OSA Continuum}, month = {May}, note = {We performed ultra-multiplex coherent anti-Stokes Raman scattering (CARS) spectroscopic imaging by using a CCD camera with a fast readout time (<1 ms). The ultra-multiplex CARS signal of a polystyrene bead was detected in the range 600–3600 cm-1 with resolution <10 cm-1. The pixel dwell time was approximately 1 ms, which was limited by the readout time of the CCD camera rather than the exposure time. CARS images of 161 × 161 pixels were obtained of the polymer beads with a total data-acquisition time of approximately 28 s even with the use of a cost-effective microchip laser source. Label-free and ultra-multiplex (18 colors) imaging of living cells was also performed with an effective exposure time of 1.8 ms based on the molecular fingerprint as well as the C-H and O-H stretching vibrational modes using a master oscillator fiber amplifier laser source.}, title = {Ultra-multiplex CARS spectroscopic imaging with 1-millisecond pixel dwell time}, volume = {2}, year = {2019}, yomi = {カノウ, ヒデアキ and ノグチ, マサユキ} }