@article{oai:tsukuba.repo.nii.ac.jp:00031416, author = {玉田, 大輝 and 巨瀬, 勝美 and Tamada, D. and Kose, K.}, issue = {9}, journal = {IEEE Transactions on Medical Imaging}, month = {Sep}, note = {A compressed sensing method using a cross sampling and self-calibrated off-resonance correction is proposed. Estimation of the magnetic field inhomogeneity based on image registration enables the off-resonance correction with no additional radio-frequency pulses or acquisitions. In addition to this advantage, a fast and straightforward calculation was achieved by using the first-order components of the magnetic field inhomogeneity. Imaging experiments using a phantom and a chemically fixed mouse demonstrated practical benefits in improving blurring and artifacts in magnetic resonance images in low field magnetic resonance imaging systems.}, pages = {1905--1912}, title = {Two-Dimensional Compressed Sensing Using the Cross-sampling Approach for Low-Field MRI Systems}, volume = {33}, year = {2014} }